The scope of the Accredited Calibration Laboratory Certificate No. AP 108

Type of instrument | Measuring range | Calibration and Measurement Capability (CMC) |
Electrical simulation of physical quantities | ||
– temperature simulators type: R, S, B type: J, K, N, T, E type: RTD | (-30 ÷ 0) °C (0 ÷ 1820) °C (-270 ÷ 1370) °C (-200 ÷ 400) °C (400 ÷ 850) °C | 0,6 °C 0,5 °C 0,3 °C 0,025 °C 0,09 °C |
– temperature indicators (meters) (including temperature controllers) – temperature recorders type: R, S, B type: J, K, N, T, E type: RTD | (-30 ÷ 0) °C (0 ÷ 1000) °C (1000 ÷ 1820) °C (-270 ÷ 500) °C (500 ÷ 1370) °C (-200 ÷ 0) °C (0 ÷ 400) °C (400 ÷ 850) °C | 1,2 °C 0,7 °C 1,0 °C 0,35 °C 0,60 °C 0,13 °C 0,09 °C 0,60 °C |
– temperature transmitters type: R, S, B type: J, K, N, T, E type: RTD | (0 ÷ 500) °C (500 ÷ 1760) °C (-180 ÷ 1370) °C (-200 ÷ 0) °C (0 ÷ 400) °C (400 ÷ 850) °C | 0,7 °C 0,5 °C 0,35 °C 0,08 °C 0,13 °C 0,15 °C |
Electrical thermometer | ||
– thermocouples made of noble and non-precious metals | (-30 ÷ 240) °C (240 ÷ 1085) °C (1085 ÷ 1200) °C (0 ÷ 250) °C (250 ÷ 600) °C (600 ÷ 1000) °C | 0,4 °C 0,7 °C 1,0 °C 0,8 °C 1,0 °C 1,2 °C |
– resistance temperature detectors (RTDs) | (-30 ÷ 100) °C (100 ÷ 200) °C (200 ÷ 240) °C (0 ÷ 250) °C (250 ÷ 400) °C (400 ÷ 600) °C | 0,04 °C 0,05 °C 0,06 °C 0,3 °C 0,5 °C 0,7 °C |
– temperature transmitters (equipped with temperature sensors) | (-30 ÷ 240) °C (240 ÷ 1200) °C 0 ÷ 250) °C (250 ÷ 600) °C (600 ÷ 1000) °C | 0,06 °C 0,8 °C 0,5 °C 0,8 °C 1,2 °C |
– electrical thermometers (including electronic) | (-30 ÷ 240) °C (240 ÷ 1200) °C (0 ÷ 250) °C (250 ÷ 400) °C (400 ÷ 600) °C (600 ÷ 1000) °C | 0,06 °C 0,8 °C 0,3 °C 0,5 °C 0,9 °C 1,1 °C |
– electrical thermometers (with temperature recording) | (-30 ÷ 240) °C (240 ÷ 450) °C (450 ÷ 1100) °C (1100 ÷ 1200) °C | 0,06 °C 0,5 °C 0,8 °C 1,7 °C |
– electrical (including electronic thermometers, with temperature recording) | (0 ÷ 250) °C (250 ÷ 400) °C (400 ÷ 600) °C (600 ÷ 1000) °C | 0,3 °C 0,5 °C 0,9 °C 1,1 °C |
– temperature calibrators | (0 ÷ 300) °C (300 ÷ 1000) °C | 0,5 °C 1,5 °C |
Non-electrical thermometry | ||
– dial thermometers (bimetallic, manometric, dilatation) | (-30 ÷ 50) °C (50 ÷ 400) °C (400 ÷ 600) °C (0 ÷ 250) °C (250 ÷ 600) °C | 0,5 °C 2,0 °C 4,0 °C 2,0 °C 4,0 °C |
The CMC uncertainty is the expanded uncertainty with a coverage probability of approximately 95 % and is expressed in units of the measurand.